Steroids 2015-07-01

Detection of dihydroxycholesterols in human plasma using HPLC-ESI-MS/MS.

Ratna Karuna, Isabelle Christen, Andreas W Sailer, Francis Bitsch, Juan Zhang

Index: Steroids 99 , 131-8, (2015)

Full Text: HTML

Abstract

We report a straightforward sample preparation procedure and a direct liquid chromatography electrospray ionization tandem mass spectrometry (LC-ESI-MS/MS) method for the analysis of 7alpha,25-dihydroxycholesterol (7α25-OHC) and 7alpha,27-dihydroxycholesterol (7α27-OHC). By applying a slow protein precipitation approach using cold ethanol, we were able to detect and quantify 7α25-OHC and 7α27-OHC in a fast and reliable manner. The average concentrations from 20 healthy individuals were determined to be 0.21±0.05nM for 7α25-OHC and 3.4±0.1nM for 7α27-OHC. In addition, we are the first to report the average degrees of esterification (n=8) to be 73.8% and 82% for 7α25-OHC and 7α27-OHC, respectively. Using the established method, we achieved the sensitivity sufficient for detecting low abundant dihydroxylated oxysterols in healthy individuals. This result should enable extension of these studies towards a comprehensive analysis of oxysterol levels under disease conditions. Copyright © 2015 Elsevier Inc. All rights reserved.

Related Compounds

Structure Name/CAS No. Articles
Methanol Structure Methanol
CAS:67-56-1
Formic Acid Structure Formic Acid
CAS:64-18-6
Butylated hydroxytoluene Structure Butylated hydroxytoluene
CAS:128-37-0
(3β,7α)-Cholest-5-ene-3,7,25-triol Structure (3β,7α)-Cholest-5-ene-3,7,25-triol
CAS:64907-22-8
25-Hydroxycholesterol Structure 25-Hydroxycholesterol
CAS:2140-46-7