Journal of chromatography. A 2014-12-05

Effect of injection matrix concentration on peak shape and separation efficiency in ion chromatography.

Ya Zhang, Charles A Lucy

文献索引:J. Chromatogr. A. 1371 , 177-83, (2014)

全文:HTML全文

摘要

In HPLC, injection of solvents that differ from the eluent can result in peak broadening due to solvent strength mismatch or viscous fingering. Broadened, distorted or even split analyte peaks may result. Past studies of this injection-induced peak distortion in reversed phase (RPLC) and hydrophilic interaction (HILIC) liquid chromatography have led to the conclusion that the sample should be injected in the eluent or a weaker solvent. However, there have been no studies of injection-induced peak distortion in ion chromatography (IC). To address this, injection-induced effects were studied for six inorganic anions (F-, Cl-, NO2-, Br-, NO3- and SO4(2-)) on a Dionex AS23 IC column using a HCO3-/CO3(2-) eluent. The VanMiddlesworth-Dorsey injection sensitivity parameter (s) showed that IC of anions has much greater tolerance to the injection matrix (HCO3-/CO3(2-) herein) mismatch than RPLC or HILIC. Even when the injection contained a ten-fold greater concentration of HCO3-/CO3(2-) than the eluent, the peak shapes and separation efficiencies of six analyte ions did not change significantly. At more than ten-fold greater matrix concentrations, analyte anions that elute near the system peak of the matrix were distorted, and in the extreme cases exhibited a small secondary peak on the analyte peak front. These studies better guide the degree of dilution needed prior to IC analysis of anions.Copyright © 2014 Elsevier B.V. All rights reserved.

相关化合物

结构式 名称/CAS号 全部文献
氟化钠 结构式 氟化钠
CAS:7681-49-4
纯碱 结构式 纯碱
CAS:497-19-8
甲酸 结构式 甲酸
CAS:64-18-6
氯化钠 结构式 氯化钠
CAS:7647-14-5
溴化钠 结构式 溴化钠
CAS:7647-15-6
亚硝酸钠 结构式 亚硝酸钠
CAS:7632-00-0
乙腈 结构式 乙腈
CAS:75-05-8
元明粉 结构式 元明粉
CAS:7757-82-6
邻二甲苯 结构式 邻二甲苯
CAS:95-47-6
苯甲醇 结构式 苯甲醇
CAS:100-51-6