Novel screen-printed electrode for the determination of dodecyltrimethylammonium bromide in water samples.
Gehad G Mohamed, Tamer Awad Ali, M F El-Shahat, M A Migahed, A M Al-Sabagh
Index: Drug Test. Anal. 4(12) , 1009-13, (2012)
Full Text: HTML
Abstract
The construction and electrochemical response characteristics of a screen-printed electrode (SPE) for the determination of dodecyltrimethylammonium bromide (DTAB) are described. The sensor was based on the use of DTA-tetraphenylborate ion association complex as an electroactive material in screen-printed electrode with dioctylsebacate (DOS) as a solvent mediator. In aqueous solution of pH 3, the sensor displayed a stable response for six months with reproducible potential and linear response for surfactant over the concentration range 1.20 × 10(-2) -5.6 × 10(-7) mol L(-1) at 25 °C with Nernstian slope of 55.95 ± 0.58 mV decade(-1) for detection limit of 6.8 x 10(-6) mol L(-1) . The response time was 6-10 s. The selectivity coefficients indicate excellent selectivity for DTAB over many common cations (e.g. Mg(2+), Na(+), K+, Co(2+), Ni(2+), Ca(2+), Cl(-), I(-), SO(4)(-2) and cetylpyridinium chloride (CPC). The sensor was used successfully for the determining of DTAB in pure form and water samples with average recoveries of 99.98, 98.78, and 99.99%.Copyright © 2011 John Wiley & Sons, Ltd.
Related Compounds
Related Articles:
2015-09-14
[Dalton Trans. 44 , 15135-44, (2015)]
2015-07-07
[Soft Matter 11 , 5077-86, (2015)]
2015-01-01
[Chem. Pharm. Bull. 63 , 1076-80, (2015)]
2014-09-01
[Brain 137(Pt 9) , 2493-508, (2014)]
Membrane sensors for the selective determination of donepezil hydrochloride.
2010-01-01
[J. AOAC Int. 93(2) , 549-55, (2010)]