Pharmaceutical Development and Technology 2006-02-01

Adaptation of USP types II and IV controlled release assays for sparingly soluble compounds by direct eluent HPLC analysis.

P J Missel, L E Stevens

文献索引:Pharm. Dev. Technol. 11(1) , 87-91, (2006)

全文:HTML全文

摘要

Measurement of drug release of a sparingly soluble drug by conventional methods proceeds very slowly without the aid of surfactants. Two preliminary automated methods were developed that increase sensitivity and accelerate such studies by working at very small reservoir volumes. All high pressure liquid chromatography (HPLC) equipment components were commercially available. Results are presented for the drug release of a single pellet of the sparingly soluble drug Eliprodil in two types of drug release experiments using (1) a stirred cell and (2) a flow-through method. Release rates measured from each method were comparable.

相关化合物

结构式 名称/CAS号 全部文献
依利罗地 结构式 依利罗地