Analytical evaluation of four on-site oral fluid drug testing devices.
Sylvie Vanstechelman, Cristina Isalberti, Trudy Van der Linden, Kristof Pil, Sara-Ann Legrand, Alain G Verstraete
Index: J. Anal. Toxicol. 36(2) , 136-40, (2012)
Full Text: HTML
Abstract
The use of oral fluid (OF) as an alternative matrix for the detection of drugs of abuse has increased over the last decade, leading to the need for a rapid, simple, and reliable on-site OF testing device. Four on-site OF drug testing devices (Dräger DrugTest 5000, Cozart DDS, Mavand Rapid STAT, and Innovacon OrAlert) were evaluated on 408 volunteers at drug treatment centers. UPLC-MS-MS results were used as reference to determine sensitivity, specificity and accuracy for each device, applying Belgian legal confirmation cutoffs for benzoylecgonine, cocaine, and THC (10 ng/mL); morphine and 6-acetylmorphine (5 ng/mL); and amphetamine and 3,4-methylenedioxymethylamphetamine (25 ng/mL). Sensitivity for cocaine was 50%, 50%, 27%, and 11% for DrugTest, OrAlert, Rapid STAT, and DDS 806, respectively. For opiates, sensitivities were 84%, 73%, 77%, and 65%, respectively. For THC, the sensitivities were 81%, 23%, 43%, and 28%, respectively. For amphetamines, the sensitivities were 75%, 33%, 17%, and 67%, respectively. Specificity was >88% for opiates and THC, > 90% for amphetamines, and > 97% for cocaine. All tests showed good specificity. DrugTest had the highest sensitivity, although it was still low for some analytes.
Related Compounds
Related Articles:
Formation of benzoylecgonine isopropyl ester following solid-phase extraction.
2012-03-01
[J. Anal. Toxicol. 36(2) , 141-3, (2012)]
2012-01-01
[Clin. Exp. Obstet. Gynecol. 39(1) , 36-42, (2012)]
2012-01-01
[Pharmacol. Rep. 64(6) , 1337-49, (2012)]
2012-07-01
[J. Anal. Toxicol. 36(6) , 390-8, (2012)]
Permeation and metabolism of cocaine in the nasal mucosa.
2012-12-01
[Eur. J. Drug Metab. Pharmacokinet. 37(4) , 255-62, (2012)]