Artificial Cells, Blood Substitutes and Biotechnology 2012-08-01

An amperometric biosensor for choline determination prepared from choline oxidase immobilized in polypyrrole-polyvinylsulfonate film.

Merve Özdemir, Fatma Arslan, Halit Arslan

Index: Artif. Cells Blood Substit. Immobil. Biotechnol. 40(4) , 280-4, (2012)

Full Text: HTML

Abstract

In this paper, a novel amperometric choline biosensor with immobilization of choline oxidase on electrochemically polymerized polypyrrole-polyvinylsulphonate (PPy-PVS) film has been accomplished via the entrapment technique. The effects of pH and temperature were investigated and optimum parameters were found to be 9.0 and 60 °C, respectively. There are two linear parts in the region between 1.0 × 10 (-7) - 1.0 × 10 (-6)M (R(2) = 0.997) and 1.0 × 10 (-5) - 1.0 × 10 (-3) M (R(2) = 0.986). The storage stability and operation stability of the enzyme electrode were also studied.

Related Compounds

Structure Name/CAS No. Articles
Oxidase choline Structure Oxidase choline
CAS:9028-67-5
Poly(vinylsulfonic acid, sodium salt) solution Structure Poly(vinylsulfonic acid, sodium salt) solution
CAS:9002-97-5