Prikladnaia Biokhimiia i Mikrobiologiia 2013-01-01

[Potentiometric glucose determination in human serum samples with glucose oxidase biosensor based on iodide electrode].

E Karakus, Sh Pekyardimchi, E Kilich

Index: Prikl. Biokhim. Mikrobiol. 49(2) , 209-14, (2013)

Full Text: HTML

Abstract

Glucose potentiometric biosensor was prepared by immobilizing glucose oxidase on iodide-selective electrode. The hydrogen peroxide formed after the oxidation of glucose catalysed by glucose oxidase (GOD) was oxidized by sodium molybdate (SMo) at iodide electrode in the presence ofdichlorometane. The glucose concentration was calculated from the decrease of iodide concentration determined by iodide-selective sensor. The sensitivity of glucose biosensor towards iodide ions and glucose was in the concentration ranges of 1.0 x 10(-1) - 1.0 x 10(-6) M and 1.0 x 10(-2) - 1.0 x 10(-4) M, respectively. The characterization of proposed glucose biosensor and glucose assay in human serum were also investigated.

Related Compounds

Structure Name/CAS No. Articles
Hydrogen peroxide Structure Hydrogen peroxide
CAS:7722-84-1
sodium molybdate dihydrate Structure sodium molybdate dihydrate
CAS:10102-40-6
Dichloromethane Structure Dichloromethane
CAS:75-09-2
Molybdenum Structure Molybdenum
CAS:7439-98-7
Sodium molybdate Structure Sodium molybdate
CAS:7631-95-0