前往化源商城

Talanta 2002-03-04

A biosensor for monitoring formaldehyde using a new lipophilic tetrathiafulvalene-tetracyanoquinodimethane salt and a polyurethane membrane.

R Kataky, M R Bryce, L Goldenberg, S Hayes, A Nowak

文献索引:Talanta 3rd ed., 56 , 451-458, (2002)

全文:HTML全文

摘要

A format for a disposable screen-printed biosensor was investigated for monitoring formaldehyde. The screen-printed sensor format comprised a working electrode (WE) modified with platinised carbon, a new lipophilic tetrathiafulvalene-tetracyanoquinodimethane (TTF-TCNQ) salt as mediator, a plasticised polyurethane membrane (TECOFLEX SG80, PU) with anionic sites and the enzyme (formaldehyde dehydrogenase FOH) and co-factor (beta-NAD(+)). The new lipophilic TTF-TCNQ salt gave a smooth, robust electrode surface rather than microcrystalline deposits of the salt observed with TTF-TCNQ salts. Entrapment of the enzyme and co-factor behind a plastcised PU membrane with anionic sites enhanced the sensitivity and stability of the sensor. A linear response between 10(-3) and 10(-4) mol dm(-3) was observed. To demonstrate the feasibility of the use of the screen-printed sensor to monitoring formaldehyde in ambient air, preliminary experiments were carried out in the vicinity of a local medium density fibreboard factory and its performance compared to that of a biosensor made from the same platinised carbon and enzyme system, but without the lipophilic TTF-TCNQ and membrane, using a commercial 'Rank' cell. Samples were collected using a Molspin 'Super-sniffer'.

相关化合物

结构式 名称/CAS号 全部文献
四硫富瓦烯-7,7,8,8-四氰基对醌二甲烷复合物 结构式 四硫富瓦烯-7,7,8,8-四氰基对醌二甲烷复合物
CAS:40210-84-2