前往化源商城

Journal of the American Society for Mass Spectrometry 2003-11-01

Identification of diethylene glycol monobutyl ether as a source of contamination in an ion trap mass spectrometer.

Christopher R Gibson, Cheri M Brown

文献索引:J. Am. Soc. Mass Spectrom. 14(11) , 1247-9, (2003)

全文:HTML全文

摘要

Tandem liquid chromatography-mass spectrometry coupled to online radioactive material detection (LC/RAM/MS/MS) is a technique that is used routinely for in vivo and in vitro drug metabolism studies and allows for a simultaneous correlation between radiochemical peaks and mass spectral data. The compound diethylene glycol monobutyl ether (DGBE), a component of a commercially available scintillation cocktail for RAM analysis, was identified as a source of overwhelming chemical noise in a mass spectrometer which was used in an LC/RAM/MS/MS configuration. In this report, we describe the identification of DGBE as the source of the chemical noise and the methods that were used to minimize the exposure of the mass spectrometer to volatile components of the scintillation cocktail.

相关化合物

结构式 名称/CAS号 全部文献
二乙二醇丁醚 结构式 二乙二醇丁醚
CAS:112-34-5