前往化源商城

Journal of Chromatography A 2000-01-21

High-sensitivity chromatographic detector incorporating three-dimensional charge coupled device fluorimetry.

S J Setford, S Saini

文献索引:J. Chromatogr. A. 867(1-2) , 93-104, (2000)

全文:HTML全文

摘要

This article describes a detector for liquid chromatography that provides three-dimensional, excitation-emission-intensity data of fluorescent compounds. The instrument achieves multi-wavelength excitation by employing a polychromator that disperses radiation along the vertical portion of a quartz capillary flow cell. The ensuing sample fluorescence is collected in the horizontal plane and dispersed using an emission polychromator to form a focused two-dimensional excitation-emission image at a charge coupled device (CCD) pixel array. Each vertical pixel equates to a narrowly defined excitation wavelength band, whilst horizontal pixels equate to corresponding emission wavelengths, with the intensity of light encroaching onto the CCD pixel elements providing the third-dimensional variable in the measurement. It is shown that the extreme sensitivity of CCD detection coupled to the dual polychromator arrangement makes possible the collection of fluorescence excitation-emission-intensity data of polycyclic aromatic hydrocarbons in a flow stream in less than 0.05 s with nanogram per millilitre limits of detection.

相关化合物

结构式 名称/CAS号 全部文献
苯并荧蒽 结构式 苯并荧蒽
CAS:205-82-3