前往化源商城

BioMedical Engineering OnLine 2009-01-01

Magnetic resonance flow velocity and temperature mapping of a shape memory polymer foam device.

Ward Small, Erica Gjersing, Julie L Herberg, Thomas S Wilson, Duncan J Maitland

文献索引:Biomed. Eng. Online 8 , 42, (2009)

全文:HTML全文

摘要

Interventional medical devices based on thermally responsive shape memory polymer (SMP) are under development to treat stroke victims. The goals of these catheter-delivered devices include re-establishing blood flow in occluded arteries and preventing aneurysm rupture. Because these devices alter the hemodynamics and dissipate thermal energy during the therapeutic procedure, a first step in the device development process is to investigate fluid velocity and temperature changes following device deployment.A laser-heated SMP foam device was deployed in a simplified in vitro vascular model. Magnetic resonance imaging (MRI) techniques were used to assess the fluid dynamics and thermal changes associated with device deployment.Spatial maps of the steady-state fluid velocity and temperature change inside and outside the laser-heated SMP foam device were acquired.Though non-physiological conditions were used in this initial study, the utility of MRI in the development of a thermally-activated SMP foam device has been demonstrated.

相关化合物

结构式 名称/CAS号 全部文献
四羟丙基乙二胺 结构式 四羟丙基乙二胺
CAS:102-60-3