前往化源商城

Analytica Chimica Acta 2012-02-10

Amperometric L-lactate biosensor based on screen-printed carbon electrode containing cobalt phthalocyanine, coated with lactate oxidase-mesoporous silica conjugate layer.

Takeshi Shimomura, Touru Sumiya, Masatoshi Ono, Tetsuji Ito, Taka-aki Hanaoka

文献索引:Anal. Chim. Acta 714 , 114-20, (2012)

全文:HTML全文

摘要

A novel amperometric biosensor for the measurement of L-lactate has been developed. The device comprises a screen-printed carbon electrode containing cobalt phthalocyanine (CoPC-SPCE), coated with lactate oxidase (LOD) that is immobilized in mesoporous silica (FSM8.0) using a polymer matrix of denatured polyvinyl alcohol; a Nafion layer on the electrode surface acts as a barrier to interferents. The sampling unit attached to the SPCE requires only a small sample volume of 100 μL for each measurement. The measurement of l-lactate is based on the signal produced by hydrogen peroxide, the product of the enzymatic reaction. The behavior of the biosensor, LOD-FSM8.0/Naf/CoPC-SPCE, was examined in terms of pH, applied potential, sensitivity and operational range, selectivity, and storage stability. The sensor showed an optimum response at a pH of 7.4 and an applied potential of +450 mV. The determination range and the response time for L-lactate were 18.3 μM to 1.5 mM and approximately 90s, respectively. In addition, the sensor exhibited high selectivity for L-lactate and was quite stable in storage, showing no noticeable change in its initial response after being stored for over 9 months. These results indicate that our method provides a simple, cost-effective, high-performance biosensor for l-lactate.Copyright © 2011 Elsevier B.V. All rights reserved.

相关化合物

结构式 名称/CAS号 全部文献
酞菁钴 结构式 酞菁钴
CAS:3317-67-7
乳酸氧化酶 结构式 乳酸氧化酶
CAS:9028-72-2