前往化源商城

Journal of Pharmaceutical and Biomedical Analysis 1998-08-01

A reproducible, simple, and sensitive high-performance capillary electrophoresis method for simultaneous determination of capreomycin, ofloxacin and pasiniazide in urine.

S S Zhang, H X Liu, Z B Yuan, C L Yu

文献索引:J. Pharm. Biomed. Anal. 17(4-5) , 617-22, (1998)

全文:HTML全文

摘要

Separation and determination of capreomycin (Cp), ofloxacin (Oflx) and pasiniazide (Ipa) in urine by high-performance capillary electrophoresis (HPCE) with 280 nm detection have been studied systematically. The calibration lines were linear in the range of 0.5 approximately 50 mg 1(-1), and the detection limits (S/N = 3) were 0.15, 0.20 and 0.10 mg 1(-1) for Cp, Oflx and Ipa, respectively. The recoveries for these materials from urine were higher than 93.5%. The accuracy and intra- and inter- day reproducibility of Cp, Oflx and Ipa were determined with satisfactory results. This method was successfully used for determining Cp. Oflx and Ipa in human urine.

相关化合物

结构式 名称/CAS号 全部文献
帕司烟肼 结构式 帕司烟肼