前往化源商城

Journal of Hazardous Materials 2010-04-15

Detection of explosives by positive corona discharge ion mobility spectrometry.

Mahmoud Tabrizchi, Vahideh Ilbeigi

文献索引:J. Hazard. Mater. 176(1-3) , 692-6, (2010)

全文:HTML全文

摘要

In this work, thermal decomposition has been used to detect explosives by IMS in positive polarity. Explosives including Pentaerythritol Tetranitrate (PETN), Cyclo-1,3,5-Trimethylene-2,4,6-Trinitramine (RDX), 2,4,6-Trinitrotoluene (TNT), 2,4-Dihydro-5-nitro-3H-1,2,4-triazol-3-one (NTO), 1,3,5,7-Tetranitro-1,3,5,7-tetrazocine (HMX), have been evaluated at temperatures between 150 and 250 degrees C in positive polarity in air. Explosives yield NO(x) which causes NO(+) peak to increase. Additional peaks may be used to identify the type of explosive. The limit of detection for RDX, HMX, PETN, NTO, and TNT were obtained to be 1, 10, 40, 1000, and 1000 ng, respectively.2009 Elsevier B.V. All rights reserved.

相关化合物

结构式 名称/CAS号 全部文献
戊四硝酯 结构式 戊四硝酯
CAS:78-11-5
奥克托今 结构式 奥克托今
CAS:2691-41-0