前往化源商城

Talanta 2012-11-15

Label-free impedimetric immunosensor for sensitive detection of 2,4-dichlorophenoxybutyric acid (2,4-DB) in soybean

Li Zhang, Meirong Wang, Chengyin Wang, Xiaoya Hu, Guoxiu Wang

文献索引:Talanta 101 , 226-32, (2012)

全文:HTML全文

摘要

Electrochemical impedance immunosensor, with its high sensitivity from electrochemical impedance analysis and ideal specificity from the immunoassay, is increasingly used in the detection of a kind of phenoxy acid herbicides which is 2,4-Dichlorophenoxybutyric acid (2,4-DB). In this experiment, synthetic 2,4-DB antibodies were immobilized on the electrode by the crosslinking of l-Cysteine/glutaraldehyde, and 2,4-DB were measured by the increase of electron-transfer resistance when the immune reaction occurred, with Fe(CN)63−/Fe(CN)64− as the probe. Under optimal conditions, the change of resistance is in a linear relationship with the logarithm of the concentration in the range of 1.0×10−7–1.0×10−3g/L (R=0.994) with the detection limit of 1.0×10−7g/L (0.1ppb). This method bears such merits as simplicity in operation, high sensitivity, wide linear range, specificity, reproducibility and good stability. The actual soybean samples were analyzed with the recovery of 82.8%–102.3%.

相关化合物

结构式 名称/CAS号 全部文献
2,4-二氯苯氧乙酸 结构式 2,4-二氯苯氧乙酸
CAS:94-75-7
2,4-二氯苯氧丁酸 结构式 2,4-二氯苯氧丁酸
CAS:94-82-6