前往化源商城

Journal of environmental health science and engineering 2014-01-01

Solid-phase microextraction fiber development for sampling and analysis of volatile organohalogen compounds in air.

Seyed Ghavameddin Attari, Abdolrahman Bahrami, Farshid Ghorbani Shahna, Mahmoud Heidari

文献索引:J. Environ. Health. Sci. Eng. 12(1) , 123, (2014)

全文:HTML全文

摘要

A green, environmental friendly and sensitive method for determination of volatile organohalogen compounds was described in this paper. The method is based on a homemade sol-gel single-walled carbon nanotube/silica composite coated solid-phase microextraction to develop for sampling and analysis of Carbon tetrachloride, Benzotrichloride, Chloromethyl methyl ether and Trichloroethylene in air. Application of this method was investigated under different laboratory conditions. Predetermined concentrations of each analytes were prepared in a home-made standard chamber and the influences of experimental parameters such as temperature, humidity, extraction time, storage time, desorption temperature, desorption time and the sorbent performance were investigated. Under optimal conditions, the use of single-walled carbon nanotube/silica composite fiber showed good performance, high sensitive and fast sampling of volatile organohalogen compounds from air. For linearity test the regression correlation coefficient was more than 98% for analyte of interest and linear dynamic range for the proposed fiber and the applied Gas Chromatography-Flame Ionization Detector technique was from 1 to 100 ngmL(-1). Method detection limits ranged between 0.09 to 0.2 ngmL(-1) and method quantification limits were between 0.25 and 0.7 ngmL(-1). Single-walled carbon nanotube/silica composite fiber was highly reproducible, relative standard deviations were between 4.3 to 11.7 percent.

相关化合物

结构式 名称/CAS号 全部文献
三氟乙酸(TFA) 结构式 三氟乙酸(TFA)
CAS:76-05-1
三氯乙烯 结构式 三氯乙烯
CAS:79-01-6
三氯甲苯 结构式 三氯甲苯
CAS:98-07-7
氯甲基甲醚 结构式 氯甲基甲醚
CAS:107-30-2