前往化源商城

American Journal of Physiology - Cell Physiology 2013-04-01

Redox artifacts in electrophysiological recordings.

Jonathan M Berman, Mouhamed S Awayda

文献索引:Am. J. Physiol. Cell Physiol. 304(7) , C604-13, (2013)

全文:HTML全文

摘要

Electrophysiological techniques make use of Ag/AgCl electrodes that are in direct contact with cells or bath. In the bath, electrodes are exposed to numerous experimental conditions and chemical reagents that can modify electrode voltage. We examined voltage offsets created in Ag/AgCl electrodes by exposure to redox reagents used in electrophysiological studies. Voltage offsets were measured in reference to an electrode separated from the solution by an agar bridge. The reducing reagents Tris-2-carboxyethly-phosphine, dithiothreitol (DTT), and glutathione, as well as the oxidizing agent H(2)O(2) used at experimentally relevant concentrations reacted with Ag in the electrodes to produce voltage offsets. Chloride ions and strong acids and bases produced offsets at millimolar concentrations. Electrolytic depletion of the AgCl layer, to replicate voltage clamp and sustained use, resulted in increased sensitivity to flow and DTT. Offsets were sensitive to electrode silver purity and to the amount and method of chloride deposition. For example, exposure to 10 μM DTT produced a voltage offset between 10 and 284 mV depending on the chloride deposition method. Currents generated by these offsets are significant and dependent on membrane conductance and by extension the expression of ion channels and may therefore appear to be biological in origin. These data demonstrate a new source of artifacts in electrophysiological recordings that can affect measurements obtained from a variety of experimental techniques from patch clamp to two-electrode voltage clamp.

相关化合物

结构式 名称/CAS号 全部文献
氯化银 结构式 氯化银
CAS:7783-90-6
DL-二硫苏糖醇 结构式 DL-二硫苏糖醇
CAS:3483-12-3
氯化铵 结构式 氯化铵
CAS:12125-02-9