前往化源商城

Analytical Biochemistry 1987-02-01

Correction for contaminant fluorescence in frequency-domain fluorometry.

J R Lakowicz, R Jayaweera, N Joshi, I Gryczynski

文献索引:Anal. Biochem. 160(2) , 471-9, (1987)

全文:HTML全文

摘要

We describe a general method to correct for contaminant fluorescence when using the technique of frequency-domain fluorometry. The method can be applied regardless of the origin of the background signal, from scattered light, impurity fluorescence, or both. The procedure requires measurement of the frequency-dependent phase and modulation of the background at enough frequencies to approximate the decay law of the background. We also describe a general method to propagate the uncertainties in the measured phase and modulation values into the corrected values. This propagation is necessary to ensure proper weighting of the frequency-dependent data in the least-squares fitting algorithms. The practical usefulness of this correction method is demonstrated using frequency-domain data for one and two component mixtures which were deliberately contaminated with scattered light and/or other fluorophores.

相关化合物

结构式 名称/CAS号 全部文献
9,10-二苯基蒽 结构式 9,10-二苯基蒽
CAS:1499-10-1
1,4-双(5-苯基-2-噁唑基)苯 结构式 1,4-双(5-苯基-2-噁唑基)苯
CAS:1806-34-4