前往化源商城

American Journal of Kidney Diseases 2013-01-01

Thrombocytopenia associated with one type of polysulfone hemodialysis membrane: a report of 5 cases.

Luyi De Prada, Jean Lee, Avrum Gillespie, Joseph Benjamin

文献索引:Am. J. Kidney Dis. 61(1) , 131-3, (2013)

全文:HTML全文

摘要

Much progress has been made in the biocompatibility of hemodialysis membranes with the development of synthetic membranes. However, despite better biocompatibility, thrombocytopenia has been observed in patients who are dialyzed with various polysulfone membranes. Recent publications have attributed this phenomenon to the electron-beam sterilization of dialyzers. We report 5 cases of thrombocytopenia in patients treated by hemodialysis with an electron-beam sterilized polysulfone membrane. The thrombocytopenia resolved after switching to either of 2 polysulfone membranes, one of which also was electron-beam sterilized. Although we were unable to elucidate the exact biochemical mechanism, the cause of the thrombocytopenia in this series appears to be related to the type of polysulfone and not due to the sterilization technique.Copyright © 2012 National Kidney Foundation, Inc. Published by Elsevier Inc. All rights reserved.

相关化合物

结构式 名称/CAS号 全部文献
聚(氧代-1,4-亚苯基磺酰-1,4-苯撑) 结构式 聚(氧代-1,4-亚苯基磺酰-1,4-苯撑)
CAS:25608-63-3
聚砜树脂 结构式 聚砜树脂
CAS:25135-51-7