Estimation, prevention, and quality control of carbon dioxide loss during aerobic sample processing.
Z L Bandi
Index: Clin. Chem. 27(10) , 1676-81, (1981)
Full Text: HTML
Abstract
We find that 2 to 6 mmol of carbon dioxide per liter (mean: 4.1 mmol/L) is lost during routine laboratory processing of patients' serum samples after centrifugation. Additional CO2 may be lost if evacuated blood-collection tubes are not filled completely during phlebotomy. More than 2 mmol of CO2 per liter is lost from samples stored in tightly stoppered tubes for 120 min if the tubes are less than half full. In extreme cases, 8 mmol of CO2 per liter may be lost from samples exposed to room air in open cups of automated micro-sample instruments. Clinically significant CO2 loss (greater than 2 mmol/L) before analysis is not detected by many laboratories because the generally accepted quality-control programs monitor only the very last step of the analytical process. A valid CO2 quality-control program should include samples with high as well as the generally used low pCO2 values. Alkalinization of serum and plasma samples with tert-butylamine prevents CO2 loss. Optimum tert-butylamine concentration, pH, and pCO2 were about 14 to 16 mmol/L, 9.0 to 9.3, and 0.4 to 1.5 mmHg (about 50 to 200 Pa).
Related Compounds
Related Articles:
Multilayered Thin Films from Boronic Acid-Functional Poly(amido amine)s.
2015-09-01
[Pharm. Res. 32(9) , 3066-86, (2015)]
Ion chromatography of amylamine and tert.-butylamine in pharmaceuticals.
1996-07-19
[J. Chromatogr. A. 739(1-2) , 343-9, (1996)]
[Hygienic standard for tert-butylamine in water].
1985-12-01
[Gig. Sanit. (12) , 70-1, (1985)]
1980-01-01
[Biochim. Biophys. Acta 613(1) , 140-52, (1980)]
2010-08-21
[Org. Biomol. Chem. 8(16) , 3631-4, (2010)]