Journal of the American Society for Mass Spectrometry 2016-07-01

Direct and Sensitive Detection of CWA Simulants by Active Capillary Plasma Ionization Coupled to a Handheld Ion Trap Mass Spectrometer.

Jan-Christoph Wolf, Raphael Etter, Martin Schaer, Peter Siegenthaler, Renato Zenobi

Index: J. Am. Soc. Mass Spectrom. 27 , 1197-202, (2016)

Full Text: HTML

Abstract

An active capillary plasma ionization (ACI) source was coupled to a handheld mass spectrometer (Mini 10.5; Aston Labs, West Lafayette, IN, USA) and applied to the direct gas-phase detection and quantification of chemical warfare agent (CWA) related chemicals. Complementing the discontinuous atmospheric pressure interface (DAPI) of the Mini 10.5 mass spectrometer with an additional membrane pump, a quasi-continuous sample introduction through the ACI source was achieved. Nerve agent simulants (three dialkyl alkylphosphonates, a dialkyl phosporamidate, and the pesticide dichlorvos) were detected at low gas-phase concentrations with limits of detection ranging from 1.0 μg/m(3) to 6.3 μg/m(3). Our results demonstrate a sensitivity enhancement for portable MS-instrumentation by using an ACI source, enabling direct, quantitative measurements of volatile organic compounds. Due to its high sensitivity, selectivity, low power consumption (<80 W) and weight (<13 kg), this instrumentation has the potential for direct on-site CWA detection as required by military or civil protection. Graphical Abstract ᅟ.


Related Compounds

Related Articles:

Detection of nerve agent stimulants based on photoluminescent porous silicon interferometer.

2012-01-01

[Nanoscale Res. Lett. 7 , 527, (2012)]

Solid-phase microextraction low temperature plasma mass spectrometry for the direct and rapid analysis of chemical warfare simulants in complex mixtures.

2016-06-21

[Analyst 141 , 3714-21, (2016)]