Analysis of tofisopam in human serum by column-switching semi-micro high-performance liquid chromatography and evaluation of tofisopam bioequivalency.
Soo Kyoung Baek, Sung Jin Choi, Jeong Soo Kim, Eun Jeon Park, Dong Hwan Sohn, Hee-Yong Lee, Hye Suk Lee
Index: Biomed. Chromatogr. 16(4) , 277-81, (2002)
Full Text: HTML
Abstract
A rapid and sensitive column-switching semi-micro HPLC method is described for the direct analysis of tofisopam in human serum. The sample (100 microL) was directly injected onto the precolumn (Capcell Pak MF Ph-1), where unretained proteins were eluted to waste. Tofisopam was then eluted into an enrichment column using 13% acetonitrile in 50 mM phosphate buffer (pH 7.0) containing 5 mM sodium octanesulfonate and subsequently into the analytical column using 43% acetonitrile in 0.1% phosphoric acid containing 5 mM sodium octanesulfonate. The detection limit (2 ng/mL), good precision (CV < or = 4.2%) and speed (total analysis time 24 min) of the present method were sufficient for drug monitoring. This method was successfully applied to a bioequivalence test of two commercial tofisopam tablets.Copyright 2002 John Wiley & Sons, Ltd.
Related Compounds
Related Articles:
2010-11-01
[J. Neural Transm. Gen. Sect. 117(11) , 1319-25, (2010)]
2013-01-01
[Eksp. Klin. Farmakol. 76(4) , 15-7, (2013)]
2008-01-15
[Aliment. Pharmacol. Ther. 27(2) , 197-206, (2008)]
2007-05-15
[Toxicol. Appl. Pharmacol. 221(1) , 111-8, (2007)]
2006-09-29