Journal of Chromatography B 2010-09-01

An optimized and validated RP-HPLC/UV detection method for simultaneous determination of all-trans-retinol (vitamin A) and alpha-tocopherol (vitamin E) in human serum: comparison of different particulate reversed-phase HPLC columns.

Abad Khan, Muhammad I Khan, Zafar Iqbal, Yasar Shah, Lateef Ahmad, David G Watson

Index: J. Chromatogr. B. Analyt. Technol. Biomed. Life Sci. 878(25) , 2339-47, (2010)

Full Text: HTML

Abstract

A novel, simple and fast reversed-phase HPLC/UV method was developed, optimized for various chromatographic conditions, and validated according to international guidelines for simultaneous determination of all-trans-retinol and alpha-tocopherol in human serum using retinyl acetate as internal standard in the concentration of 0.5 microg/ml. A liquid-phase extraction was applied to the 250 microl of serum with n-hexane-dichloromethane mixture (70:30, v/v), in two steps, using ethanol-methanol mixture (95:5, v/v) for protein precipitation and BHT (butylated hydroxy toluene) as stabilizer for sample preparation. Both analytes were analyzed on Kromasil 100 C(18) column (150 mm x 4.6 mm, 5 microm), Brownlee analytical (Perkin Elmer) C(18) column (150 mm x 4.6 mm, 5 microm), and Supelco (Supelcosil) LC-18 column (150 mm x 3 mm, 3 microm), protected by a Perkin Elmer C(18) (30 mm x 4.6 mm, 10 microm; Norwalk, USA) pre-column guard cartridge, at 292 nm wavelength, using methanol-water (99:1, v/v), in isocratic mode as mobile phase applied at flow rate of 1.5 ml/min and 1 ml/min for both 5 microm and 3 microm columns, respectively. Complete separation of all the analytes was achieved in 3 and 6 min on 3 microm and 5 microm columns, respectively by injecting 20 microl of sample into the HPLC system by autosampler, keeping column oven temperature at 25 degrees C. Different particulate reversed-phase chromatographic columns were evaluated in order to select the best column in terms of sensitivity, selectivity, resolution and short run time of both the analytes and it was concluded that 3 microm columns are better to be used in clinical set up as well as in laboratories for the separation of these analytes in a shorter time as compared with 5 microm columns. The method was validated and applied for the analysis of all-trans-retinol and alpha-tocopherol in the serum of human volunteers.2010 Elsevier B.V. All rights reserved.


Related Compounds

Related Articles:

Simultaneous analysis for water- and fat-soluble vitamins by a novel single chromatography technique unifying supercritical fluid chromatography and liquid chromatography.

2014-10-03

[J. Chromatogr. A. 1362 , 270-7, (2014)]

Effects of microbial phytase on apparent and standardized total tract digestibility of calcium in calcium supplements fed to growing pigs.

2015-05-01

[J. Anim. Sci. 93 , 2255-64, (2015)]

Ultra-performance liquid chromatographic determination of tocopherols and retinol in human plasma.

2014-10-01

[J. Chromatogr. Sci. 52(9) , 1065-70, (2014)]

DEVTA: results from the biggest clinical trial ever.

2013-04-27

[Lancet 381(9876) , 1439-41, (2013)]

A randomized controlled trial to assess the impact of dietary energy sources, feed supplements, and the presence of super-shedders on the detection of Escherichia coli O157:H7 in feedlot cattle using different diagnostic procedures.

2010-09-01

[Foodborne Pathog. Dis. 7(9) , 1071-81, (2010)]

More Articles...