A low dielectric constant customized phantom design to measure RF coil nonuniformity.
P S Tofts, G J Barker, T L Dean, H Gallagher, A P Gregory, R N Clarke
Index: Magn. Reson. Imaging 15(1) , 69-75, (1997)
Full Text: HTML
Abstract
The selection of a suitable low dielectric constant liquid for radiofrequency coil nonuniformity measurements is described. Measurements of dielectric constant (relative permittivity) were made on a range of candidate liquids. After excluding liquids that were too inflammable or too viscous, Esso Bayol 82 oil (dielectric constant epsilon' = 2.37) was chosen. At 1.5 T, a 27 cm diameter cylindrical phantom (test object) filled with Bayol 82 has a maximum nonuniformity of 1.9% arising from radiofrequency standing waves. The maximum diameter cylinder that can be used without the nonuniformity exceeding 2% is given for a range of liquids and field strengths. The construction of customized containers that fit closely inside a radiofrequency head coil from glass fiber reinforced resin ("fiberglass") is described. Thermal expansion of the liquid takes place without a rise in the internal pressure in the container.