Annals of Clinical Biochemistry 1983-05-01

Optical methods for monitoring temperature in spectrophotometric analysers.

T D O'Leary, J L Badenoch, R Bais

Index: Ann. Clin. Biochem. 20(Pt 3) , 153-6, (1983)

Full Text: HTML

Abstract

A procedure is described for monitoring the temperature in the reaction cuvettes of analytical systems that use photometers. The method employs the temperature-dependent change in absorbance of solutions of either 3,5-dinitrosalicylic acid or cresol red. The procedure is simple to perform and is especially useful in monitoring the temperature in instruments such as centrifugal analysers where the reaction cuvette is inaccessible. In some of the instruments studied, methodological changes were required to ensure that reactions were carried out at the selected temperature.

Related Compounds

Structure Name/CAS No. Articles
Cresol red Structure Cresol red
CAS:1733-12-6
3,5-Dinitro-2-hydroxybenzoic acid Structure 3,5-Dinitro-2-hydroxybenzoic acid
CAS:609-99-4
Cresol Red Structure Cresol Red
CAS:62625-29-0