J E Van Nuwenborg, D Stöckl, L M Thienpont
Index: Eur. J. Clin. Chem. Clin. Biochem. 35(4) , 297-300, (1997)
Full Text: HTML
We evaluated bias and inaccuracy of four frequently used routine test systems for serum total calcium, using a candidate reference method based on ion chromatography. The mean biases and 95% confidence intervals that we observed were 0.0 +/- 0.59% for Johnson & Johnson arsenazo(III), 1.3 +/- 0.62% for Beckman arsenazo(III), -0.4 +/- 0.44% for Beckman applying ion selective electrode measurement after sample dilution, and -1.9% +/- 0.42% for Boehringer o-cresolphthalein. The inaccuracy of all test systems was usually < 4.7% (calculated as deviation of singlicates from ion chromatography). Both bias and inaccuracy are discussed in the light of specifications set by expert groups or derived from the biological variation of serum total calcium. The study revealed that the intrinsic quality of commonly used test systems for serum total calcium satisfies even some of the more stringent criteria for method bias and inaccuracy.
Structure | Name/CAS No. | Molecular Formula | Articles |
---|---|---|---|
![]() |
o-cresolphthalein
CAS:596-27-0 |
C22H18O4 |
Hepatorenal protective effect of Antistax(®) against chemica...
2015-05-01 [Pharmacogn. Mag. 11 , S173-81, (2015)] |
Establishing pediatric reference intervals for 13 biochemica...
2014-12-01 [Clin. Biochem. 47(18) , 268-71, (2014)] |
Relative abundance of fetuin- A in peritoneal dialysis efflu...
2015-01-01 [Nephrology (Carlton.) 20(1) , 6-10, (2014)] |
Increase in osteoclastogenesis in an obese Otsuka Long-Evans...
2015-09-01 [Mol. Med. Report. 12 , 3874-80, (2015)] |
Decline in calcitonin receptor expression in osteocytes with...
2014-05-01 [J. Endocrinol. 221(2) , 181-91, (2014)] |
Home | MSDS/SDS Database Search | Journals | Product Classification | Biologically Active Compounds | Selling Leads | About Us | Disclaimer
Copyright © 2024 ChemSrc All Rights Reserved