Journal of Chromatographic Science 2015-02-01

An optimized method for corticosterone analysis in mouse plasma by ultra-performance liquid chromatography-full-scan high-resolution accurate mass spectrometry.

Hui Li, Li-Ping Sheng, Bing Wang, Zong-Lin Yang, Shu-Ying Liu

Index: J. Chromatogr. Sci. 53(2) , 285-94, (2015)

Full Text: HTML

Abstract

A simple method based on liquid chromatography-full-scan high-resolution accurate mass spectrometry (LC-HRMS) using a quadrupole time-of-flight (TOF) mass spectrometry was developed and optimized for corticosterone quantification in mouse plasma. Mouse plasma (100 μL) was extracted with methyl tert-butyl ether using prednisone as internal standard. Separation was performed on a short C18 column using a methanol-water gradient. Full-scan data were acquired in the TOF only mode, and extracted ion chromatograms were generated post-acquisition with the extract masses of the analytes. Enhanced sensitivity and reproducibility were acquired with optimized mass parameters. The calibration range was 8.24-412 ng/mL, and the limit of quantitation was 5.088 g/mL. Accuracy was between -5.9 and 8.6%. The precision of between-run (interday) and within-run (intraday) was within 5.6 and 6.9%, respectively. The LC-HRMS method was applied for plasma samples analysis from the stressed mice with and without ginseng treatment for the stress state estimation. © The Author 2014. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Related Compounds

Structure Name/CAS No. Articles
Methyl tert-butyl ether Structure Methyl tert-butyl ether
CAS:1634-04-4
Di-n-butyl ether suppliers in China Structure Di-n-butyl ether suppliers in China
CAS:142-96-1